Optically detected magnetic resonance (ODMR) with NV centers

Hello and welcome back! In a previous post, we introduced the electronic energy structure of the NV (technically NV^-) center. We saw that under optical excitation, there are multiple relaxation pathways, both radiative and non-radiative. The m_s=0 state was more likely to undergo the radiative relaxation pathway and emit a photon as compared to the m_s=\pm1 state. This spin-dependent fluorescence of the NV center allows for optically detected magnetic resonance (ODMR), where the relative populations of spin states can be measured via the intensity of light emission from the center.

In this post, we will discuss the basics of ODMR with NV centers. We begin by explaining the state preparation and readout, and then move on to some simple (but still very useful) experiments. Hopefully, this introduction to ODMR provides the reader with the fundamentals of the technique, and the tools to be able to understand more complex experiments that may be encountered in the literature and in research environments.

Review of spin-dependent fluorescence

As discuss in our post on the NV electronic energy structure, optically excited NV electronic states have both radiative and non-radiative relaxation pathways. The likelihood of relaxation occurring via each of these pathways from the excited triplet state depends on the m_s state of the triplet. This dependence on spin state has two significant consequences.

First, optical excitation causes the conversion of m_s = \pm1 states into m_s=0 at a greater rate than the opposite direction (m_s=0 \rightarrow m_s=\pm1). This rate difference means that we can spin polarize the NV center into the m_s=0 state via optical excitation with a laser. In reality, one does not achieve 100\% spin polarization, but the achievable polarization far exceeds thermal polarization, and thus is considered hyperpolarization (values ranging from 40-92\% have been reported).

Second, we know that the m_s=0 state is more likely to relax via a radiative pathway than the m_s=\pm1 states. The fluorescence intensity at 637 nm – the wavelength corresponding to the gap between the ground electronic state and first excited state – from the NV center thus gives information on the relative populations of the different spin state. As we shall see, ODMR will leverage both of these properties.

ODMR detection

The detected signal in ODMR is the intensity of the emitted light from the NV center after a readout laser pulse. This laser pulse creates the excited triplet state, which will then relax back into the ground state (via both radiative and non-radiative routes, with the former emitting photons at 637 nm). The m_s=0 state is “brighter” than the m_s=\pm1 states by a factor of \sim 20\%, i.e. is 20\% more likely to relax via the radiative pathway. During the readout period, the intensity of emitted light thus gives information on the relative population of the different m_s states. (Specifically, the difference in population between the m_s=0 and m_s=\pm1 states, since the m_s=-1 and m_s=+1 states are indistinguishable by their fluorescence.)

It is important to note that this measurement is destructive, as the laser pulse will effect the NV spin states and preferentially reinitialize the spins into the m_s=0 state. While this property is advantageous for creating spin hyperpolarization, here it changes the state of the system, so that after one readout pulse we have destroyed the information that we wanted to readout. This state destruction will occur on the order of < 1 \mus, a timescale during which there may not even be any photons emitted from an individual NV center. In practice the SNR of a single cycle of initialization and readout is \sim 0.1, so that one must repeat this sequence many times in order to achieve a reasonable SNR. However, with enough cycles, one is able to optically measure the relative populations of the m_s=0 and m_s=\pm1 states. Thus, we can measure the spin state of the NV center via the optical fluorescence at 637 nm! This process of measuring spin states by monitoring optical signals (i.e. emitted photons) is known as optically detected magnetic resonance.

cw-ODMR

We can consider two classes of ODMR experiments, analogous to continuous wave (cw) and pulsed experiments in standard electron paramagnetic resonance. We begin with cw-ODMR, as this experiment has a simpler setup. Note that while these experiments can be conducted at zero magnetic field or at non-zero magnetic field (ODMR on NV centers has been performed between 0 and 7 Tesla).

For a cw-ODMR experiment, we apply continuous microwaves and continuous laser illumation on our NV center. Then, we scan over different microwave frequency, and measure the fluorescence from the center. When the microwave frequency is resonant with one of the spin transitions in the center, it will induce spin transitions between m_s states. Since our center is under constant laser illumination during the experiment, we are essentially constantly initialized into the m_s=0 state. Therefore, when the microwaves are on-resonance, there will be spin transitions to the m_s=\pm1 “dark” states, and the measure fluorescence intensity will decrease. When the microwaves are off-resonance, no spin transitions are induced, and the fluorescence is unchanged.

By scanning over a range of microwave frequencies and plotting the measured fluorescence intensity vs. the frequency of applied microwaves, we are able to extract an ODMR spectrum. As opposed to a cw-EPR spectrum, where peaks appear at the frequencies corresponding to spin transitions, the ODMR spectrum will display dips in the measured fluorescence at the transition frequencies. Because of the relative ease of this cw-ODMR method compared to the pulsed alternative described below, this technique is an excellent option when the central aim is to extract a 1-D spectrum of the NV center.

Pulsed ODMR

The second class of ODMR experiments we will discuss is pulsed (or time-resolved) ODMR. Figure 1 shows the basic experimental structure. The sequence begins by initializing the NV center into the m_s=0 state with a laser pulse. A series of microwave pulses and delays then manipulate the electronic spin state. The exact sequence of operations will depend on the specific experiment, but will be designed to create changes in populations of states or coherences between states, akin to standard pulsed EPR or NMR. At then end of this sequence, a second laser pulse allows for the readout of spin state populations.

Figure 1: Experimental structure for a pulsed ODMR experiment. Laser light initializes the electrons into a m_s=0 state. Microwave pulses and delays then allow for the manipulation of the spin state state, interchanging populations and/or coherences between states. A readout pulse then causes the center to fluoresce photons at 637 nm with a spin-dependent intensity. Note that here the m_s=\pm1 states have the same energy for simplicity (i.e. we are at zero magnetic field), but that the same concepts apply at non-zero fields.

If we were to want information on nuclear spin states as well as electronic spin states, then we could add radiofrequency pulses during the middle block of the experiment to manipulate the nuclear spin states. In that case, we would also need to transfer any information about nuclear spins (coherences or population changes generated) back to our electronic spin state population for readout.

As one can imagine, there are a wide variety of different microwave/delay sequences that can be used for different forms of ODMR experiments. We will introduce the simplest possible sequence below, which establishes the fundamentals underlying many ODMR experiments. But we also encourage the interested reader to stay tuned for future blogs on more complex ODMR experiments, and to look into the literature to learn more about the wide array of different experiments in ODMR.

Ramsey Fringe method

It may still be unclear how exactly this method allows for the measurement of a useful magnetic resonance signal. In conventional NMR, we can use a single 90\degree pulse to rotate spins into the transverse plane. We then can measure the resulting precession about the external field to extract the time-domain FID. In ODMR, we can also measure a FID, but this requires two microwave pulses instead of one. This technique is known as the Ramsey Fringe method, with a basic structure shown in Figure 2.

Figure 2: Experimental procedure for ODMR FID collection via the Ramsey Fringe method. Each run has a single delay time \tau, and contributes one point to the FID. By running many experiments while varying \tau, one is able to construct the full time-domain FID.

Like most NV experiments, the sequences begins by initializing the NV center into the m_s=0 state by a laser pulse. We then apply a 90\degree(y) pulse (we will use a specific phase convention for this discussion, but other pulse phases are possible and provide similar results). This pulse forms a superposition (or coherence) between our m_s = 0 (\ket{0}) and m_s=\pm1 (\ket{1}) states,

(1)   \begin{equation*} \ket{\psi(0)} = \frac{1}{\sqrt{2}}(\ket{0} + \ket{1}) \end{equation*}

(Here we consider the case of zero field, so we can treat out NV as a two level system). Because the states \ket{0} and \ket{1} have a difference in energy \Delta E, equivalent to the zero-field splitting in our zero field case, a phase will accumulate between the two states. Thus, the state at time \tau will be

(2)   \begin{equation*} \ket{\psi(\tau)} = \frac{1}{\sqrt{2}} (\ket{0} + e^{frac{i\Delta E \tau}{\hbar}} \ket{1}) \end{equation*}

At this point, we apply a second 90\degree (y) pulse, returning out state to populations of m_s=0 and m_s=\pm1. The resulting population of the m_s=0 state will be modulated by a term cos(\frac{\Delta E \tau}{\hbar}). Thus, the transition energy \Delta E is encoded within the modulation of the m_s=0 population, which can be monitored via the fluorescence from the NV center.

It may be helpful to understand this phase accumulation and resulting population modulation by considering a couple different time points. If \tau = \frac{2 \pi \hbar}{\Delta E}, then our our state before the second pulse is

(3)   \begin{equation*} \ket{\psi(0)} = \frac{1}{\sqrt{2}}(\ket{0} + \ket{1}) \end{equation*}

which we note is identical to the state immediately following the first pulse. The second pulse will then return us to our initial state of m_s=0, and the fluorescence intensity will be at a maximum.

On the other hand, if \tau = \frac{\pi \hbar}{2 \Delta E}, then we are in the state

(4)   \begin{equation*} \ket{\psi(0)} = \frac{1}{\sqrt{2}}(\ket{0} + i \ket{1}) \end{equation*}

before the second pulse. The second pulse will then not fully recover the initial state of m_s=0, but will leave the state unchanged, so that in our final state we will have only half that initial m_s=0 population that we initially had, and our fluorescence intensity will be reduced. (Note that here we have made the assumption that we have an ensemble of NV centers, or have made the measurement many times on a single center. A single measurement on one NV center would collapse the state into either the \ket{0} or \ket{1} state, so that we could not use phrases such as “half the population”).

To generate a time domain FID, one conducts many measurements while varying \tau. As noted above, at each \tau we will have to take many measurements in order to achieve sufficient SNR in our fluorescence signal. Each value of \tau generates a single point in our FID (as compared to a standard pulsed NMR experiment, where the full FID is generated by a single pulse), so we see that this is a time consuming process. After we have generated the FID over the course of many iterations of the experiment, processing is similar to a standard FID, where a Fourier transform of the FID will yield a one-dimensional spectrum.

Conclusions

As one can imagine, more complex pulse sequences are possible, including multi-dimensional experiments, but they all build on the same principles introduced above. The spin is first optically initialized into the m_s=0 state. Pulses then convert populations into coherences, which then evolve (via delays or further microwave pulses) in time. Eventually, the coherences are returned to m_s populations for optical readout.

A similar procedure can also measure NMR signals from nuclei coupled to the NV center (such as ^{14}N or ^{13}C), but in that case both microwave and radiofrequency pulses generatre electron and nuclear spin coherences, respectively. Just as above, these coherences accumulate a transition energy dependent phase, and eventually these coherences are converted into electron spin state populations for optical readout.

You may be thinking that ODMR sounds time-consuming, and it can be! But at the same time, it offers some powerful advantages over traditional magnetic resonance techniques. Importantly, we don’t need an external field to initialize or read out spin states. (Note that we could have an external field, as ODMR works at non-zero fields by the same principles.) Additionally, one can write to and read out from a single NV center, as opposed to traditional magnetic resonance which measures an ensemble of spins. These properties make NV ODMR a powerful tool for a range of applications, from magnetometry to biomedical imaging to quantum information storage and processing. Some of these applications will be discussed in future posts, but we also encourage the interested reader to look into the wide array of literature on ODMR and NV centers to delve deeper into the state of the art in these areas.